Center for Collective Use of Scientific Equipment "Laboratory of Experimental and Applied Physics"
Two-beam scanning spectrophotometer UV-1700

Designed to measure optical transmission and absorption spectra in the wavelength range from 190 to 1100 nm.
SPECIFICATIONS
|
Spectral range |
190.0-1100.0 nm |
|
Spectral width of the slit |
1nm (190 - 1100 nm) |
|
Optical circuit |
Two-beam |
|
Accuracy of wavelength display |
Step 0.1 nm |
|
Accuracy of wavelength setting |
0.1 nm |
|
Wavelength setting error |
± 0.1 nm at 656.1 nm D2 ± 0.3 nm (190 - 1100 nm) |
|
Wavelength reproducibility |
± 0.1 nm |
|
Scan speed |
ca. 6000 nm/min |
|
Wavelength registration |
ca. 10 bis 3000 nm/min |
|
Changing the light source |
Automatic |
|
Scattered light |
Less than 0.02% (220.0 nm NaI, 340.0 nm NaNO2), Less than 1.0% (198 nm KCL) |
|
Photometric range |
Absorption: -4 ~ 4 Abs Skip: 0 ~ 400% |
|
Photometric accuracy |
± 0.002 Abs (at 0.5 Abs) ± 0.004 Abs (at 1.0 Abs) ± 0.006 Abs (at 2.0 Abs) |
|
Photometric reproducibility |
Less than ± 0.001 Abs (at 0.5 Abs) Less than ± 0.001 Abs (at 1.0 Abs) Less than ± 0.003 Abs (at 2.0 Abs) |
|
Stability |
0.001 Abs/Std max. (bei 700 nm) |
|
Deviation |
0.002 Abs max. (190 bis 200 nm) - 0.001 Abs max. (200 bis 1100 nm) |
|
Noise |
P-P 0.002 Abs max., RMS 0.0002 Abs max.* (bei 700 nm) |
|
Light source |
Halogen lamp (2000 hours), Deuterium lamp, automatic position setting and adjustment at maximum sensitivity. |
|
Detector |
Silicon photodiode |
|
Display |
6-inch LCD (320 x 240 pixel) with brightness adjustment |
|
Drift |
Less than 0.0003 Abs / h, 700 nm (after heating for 1 h) |
|
Baseline deviation |
lines ± 0.0006 Abs (190 - 1100 nm, after heating for 1 h) |
|
Noise level |
Less than 0.00005 Abs, RMS, 700 nm |
|
Memory for storing data without a PC |
Built-in (up to 24 files) + external USB drive (optional) |
|
Device control |
At the choice of the operator: from the built-in processor or from the personal computer |
|
Power |
110 V / 60 Hz or 220 V / 50 Hz |
Precision digital multimeter SIGLENT SDM3065X

SDM3065X - digital multimeter with 6-bit display and the ability to simultaneously measure two parameters.
Brief characteristics and features of the SIGLENT SDM3065X multimeter:
• bit size 6 ½
• base error (DCV): ± 0.0035%
• speed of measurement - up to 10000 measurements/ second
• measurement: taking into account the shape of the signal and distortion (True RMS), variable signal with a shift (AC + DC)
• automatic and manual selection of measurement limits
• internal data logger (10 K) with adjustable sampling rate
• Expanded sync and start menu
• 2- and 4-wire circuit for measuring resistance
• temperature measurement (thermocouples and thermoresistance / Rtd)
• mathematical functions and statistical processing (min / max/ average; dB / dBm; tolerance control; Δ-measurements, histograms, trend construction. standard deviation)
• simultaneous measurement of 2 parameters
• display of the measurement result using an analogue scale (bar-graph)
• interfaces: LAN, USB (support VXI11, USBTMC), GPIB (option - cable adapter)
• 1 GB internal memory (external USB-flash extension), file system support, built-in on-screen help (HELP)
• EasyDDM software for managing and transferring data to a computer
• color graphic TFT display (11 cm, 480 x 272 pixels)
Specifications
|
DC voltage measurement |
200 mV, 2 V, 20 V, 200 V, 1000 V |
|
DC measurement |
200 mA, 2 mA, 20 mA, 200 mA, 2 A, 10 A |
|
AC voltage measurement |
True-RMS, 200 mV, 2 V, 20 V, 200 V, 750 V |
|
AC measurement |
True-RMS, 20 mA, 200 mA, 2 A, 10 A |
|
Measurement of resistance (on the 2- and 4-wire circuit) |
200 Ohm, 2 kOhm, 20 kOhm, 200 kOhm, 2 mOhm, 10 mOhm, 100 mOhm |
|
Capacity measurement |
2 nF, 20 nF, 200 nF, 2 μF, 20 μF, 200 μF, 10,000 μF |
|
Conductivity test |
range of 2 kOhm |
|
Diode test |
Range 0 - 2 V |
|
Frequency measurement |
20 Hz ~ 1 MHz |
|
Period measurement |
1 μs ~ 0.05 s |
|
Temperature measurement |
thermocouple support, RTD temperature sensor |
|
Max. input voltage |
1000 V |
|
Interfaces |
USB Device, USB Host, LAN, GPIB |
Phosphor digital oscilloscope SIGLENT SDS1202X

SIGLENT SDS1202X is a two-channel phosphor oscilloscope with a bandwidth of 200 MHz, a sampling frequency of 1 GS/s, and a standard memory depth of up to 14 MB.
SPECIFICATIONS
|
Bandwidth |
200 MHz |
|
Channels |
2 + 1 external. |
|
Sampling frequency |
1 GS / S |
|
Max. memory depth |
14 MB / channel (one channel); 7 MB / channel (two channels) |
|
Oscillogram capture rate |
400,000 RPM (segmented sampling mode) |
|
Start types |
Edge, Slope, Pulse, Window, Runt, Interval, Time out (Dropout), Pattern, Video |
|
Vertical sensitivity |
500 µV/div - 10 V/div |
|
Mathematical operations |
addition, subtraction, multiplication, division, FFT, differentiation, integration, square root extraction |
|
FFT (processing) |
1 M points |
|
Automatic measurements |
38 types: statistics calculation, zoom operations, logical operations, mathematical operations, history, relative measurements |
|
Decoding |
I2C, SPI, UART / RS232, CAN, LIN |
|
Input / output |
USB Host, USB Device, LAN, Pass/Fail, Trigger Out, Ext Trigger |
|
Display |
7-inch (800 × 480) |
|
Power |
AC 100-240 V, 45 Hz - 440 Hz, 50 VA max. |
Precision measuring instrument LCR E4980A

Keysigt E4980A is a precision imitation meter (RLC meter) with three additional frequency options: 300 kHz, 500 kHz and 1 MHz.
SPECIFICATIONS
|
Frequency range |
from 20 Hz to 2 MHz |
|
|
Test signal level |
0 to 2 V (RMS) / 0 to 20 mA (RMS) |
|
|
Auto level control |
+ |
|
|
DC offset function |
Built-in, 1.5; 2 V ± 40 V |
|
|
DC source |
± 10 V |
|
|
Programmable sweep in list mode |
201 points |
|
|
Remote control |
GPIB |
|
|
Management via web browser |
+ |
|
|
Management commands |
Compatible with 4284A |
|
|
Basic accuracy |
0.1% ("short" mode) |
|
|
Parameters |
Cp-D/Q/G/Rp, Cs-D/Q/Rs, Lp-D/Q/G/Rp, Ls-D/Q/Rs, R-X, Z-0d/0r, G-B, Y-0d/0r Lp-Rdc , Ls-Rdc , Vdc-Idc |
|
|
Measurement time mode Model |
Standard (E4980A) |
Simplified (E4980A-005) |
|
(Short / Medium) |
330 ms / 380 ms 20 Hz |
1040 ms / 1150 ms 20 Hz |
|
100 ms / 180 ms 100 Hz |
240 ms / 380 ms 100 Hz |
|
|
20 ms / 110 ms 1 kHz |
37 ms / 200 ms 1 kHz |
|
|
7.7 ms / 92 ms 10 kHz |
25 ms / 180 ms 10 kHz |
|
|
5.7 ms / 89 ms 100 kHz |
23 ms / 180 ms 100 kHz |
|
|
5.6 ms / 88 ms 1 MHz |
23 ms / 180 ms 1 MHz |
|
|
5.6 ms / 88 ms 2 MHz |
23 ms / 180 ms 2 MHz |
|
|
Storage devices |
Internal / USB memory |
|
OWON xDS3062A digital oscilloscope

Digital oscilloscope OWON xDS3062A - multifunctional digital oscilloscope OWON, (bandwidth 60 MHz, sampling rate: 1 GV / s, memory 40 MB, 8 '' display, ADC: 12 bits).
SPECIFICATIONS
|
Bandwidth |
60 MHz |
|
Number of channels |
2 |
|
Sampling frequency |
1 GS/ s |
|
Vertical resolution |
12 bits |
|
Depth of memory |
40 MB |
|
Oscillogram recording |
45,000 oscillograms / sec |
|
Rise time (input, default) |
≤5.8 ns |
|
Horizontal scale |
2 ns / div- 1000 s / div |
|
Input impedance |
1 mOhm ± 2%, in parallel with 15 pF ± 5 pF |
|
Max. input voltage |
1 mOhm ≤ 300 Vskz; |
|
Measurement error |
±1% |
|
DC error |
average ≥16 : ± ( 3% +0.05 div) for △ V |
|
Correlation by probe coefficient |
0.001X - 1000X, steps 1 - 2 - 5 |
|
Woofer feedback (ace, -3 dB) |
≥5 Hz |
|
Interpolation |
(sinx)/x,x |
|
Login modes |
DC, AC, GND |
|
Vertical sensitivity |
1 mV / div- 10 V / div(at input) |
|
Start types |
Edge, Video, Pulse, Slope, Runt, Windows, Timeout, Nth Edge, Logic, I2C, SPI, RS232, CAN (optional) |
|
Bus decoding |
I2C, SPI, RS232, CAN |
|
Start modes |
Auto, Normal, Single |
|
Vertical range |
± 2V (1mV / div~ 50mV / div) ; ± 20V (100mV / div~ 1V / div) ; ± 200V (2V / div~ 10V / div) |
|
Mathematical functions |
|
|
Line / field frequency (video) |
NTSC, PAL and SECAM standards |
|
Cursor measurements |
△ V, and △ T between cursors, △ V and △ T between cursors, and autocursors |
|
Automatic measurements |
Vpp, Vavg, Vrms, Freq, Period, Week RMS, Cursor RMS, Vmax, Vmin, Vtop, Vbase, Vamp, Overshoot, Phase A→B ↑, Phase A→B↓, Preshoot, Rise Time, Fall Time,+Width, -Width, +Duty, -Duty, Duty Cycle, Delay A→B ↑, Delay A→B↓, +Pulse Count, -Pulse Count, Rise Edge Count, Fall Edges Count, Area, Cycle Area |
|
Storage of oscillograms |
100 oscillograms |
|
Interfaces |
USB host, USB device, Trig Out (P / F), LAN, VGA (optional) |
|
Display |
8 "color LCD, 800 x 600 pixels |
|
Power |
100 V - 240 V AC, 50/60 Hz, CAT II |
|
Battery |
3.7 V, 13200 mA |
|
Dimensions |
340 × 177 × 90 mm |
SIGLENT SDG6022X signal generator

The SIGLENT SDG6022X signal generator is a device used to generate high-quality signals and is characterized by a bandwidth of 200 MHz, a maximum sampling rate of 2.4 GS/s and a vertical resolution of 16 bits.
SPECIFICATIONS
|
Bandwidth |
200 MHz |
|
Number of channels |
2 |
|
Sampling frequency |
2.4 GW / s (2X interpolation) |
|
Vertical resolution |
16 bit |
|
The length of an arbitrary signal |
2 ~ 20 MB |
|
Frequency meter |
up to 400 MHz |
|
Max. Amplitude |
20 Vpp <40 MHz |
|
Interface |
Standard: USB Host, USB Device, LAN |
|
Display |
Touch 4.3 ”display, 480 × 272 RGB |
Spectral ellipsometer Smart SE

Smart SE ellipsometer is a multifunctional spectral ellipsometer for fast and accurate analysis of thin films, which allows determining film thicknesses from a few angstroms to 20 microns, optical characteristics (n, k), roughness, gradient and anisotropic layers and much more.
SPECIFICATIONS
|
Spectral range, nm |
450 – 1000 |
|
Spectral resolution, nm |
better than 3 |
|
Light source |
combined |
|
Measurement time, sec |
<1 to 10 |
|
The size of the analytical spot, microns |
75 x 150; 100 x 250; 100 x 500; |
|
Angle of incidence, deg |
from 45 to 90 with step 5 |
|
Sample size, mm |
up to 200 |
|
Adjusting of the sample position |
manual, 17 mm in height and inclination |
|
Measurement quality: |
|
|
Accuracy (directly through the county) |
Ψ=450±0,050; Δ=00±0,20 |
|
Accuracy (thickness on oxide 1000 angstroms) |
0,04% |
|
Reproducibility (thickness on the oxide 1000 angstroms) |
±0,02% |
|
Dimensions (LxWxH), cm |
100 x 46 x 23 |
OWON digital precision multimeter XDM3041

XDM3041 is a professional OWON digital multimeter with statistical processing functions and an advanced set of interfaces (LAN, RS232, USB). The XDS3041 multimeter has excellent measurement accuracy (base error: 0.02%), designed to measure AC / DC voltage, AC / DC current, resistance, capacitance, frequency / period, temperature, diode testing and conductivity testing. Resistance measurement is performed on both 2-wire and 4-wire circuit, which eliminates the resistance of the probes.
SPECIFICATIONS
|
Specifications |
Range |
Frequency |
Basic error +% of range |
|||
|
DC voltage |
600 mV |
|
0,02%+0,01% |
|||
|
0,02%+0,01% |
||||||
|
0,02%+0,01% |
||||||
|
0,02%+0,01% |
||||||
|
AC voltage (TrueRMS) |
600 mV |
20 Hz - 45 Hz |
2,0%+0,10% |
|||
|
45 Hz - 20 kHz |
0,2%+0,06% |
|||||
|
20 kHz - 50 kHz |
1,0%+0,05% |
|||||
|
50 kHz - 100 kHz |
3,0%+0,08% |
|||||
|
DC current |
600.00 μA |
|
0,06%+0,02% |
|||
|
AC current (TrueRMS ) |
60,000 mA |
20 Hz - 45 Hz |
2,0%+0,10% |
|||
|
45 Hz-2 kHz |
0,50%+0,10% |
|||||
|
2 kHz-10 kHz |
2,5%+0,20% |
|||||
|
Resistance |
600.00 Ohm |
|
0,040%+0,01% |
|||
|
Frequency ( 200mV-750V ) |
20 Hz - 2 kHz |
|
0,01%+0,003% |
|||
|
Frequency (20mA-10V) |
20 Hz - 2 kHz |
|
0,01%+0,003% |
|||
|
Capacity |
2,000 nF |
200 nA |
3,0%+1,0% |
|||
SIGLENT SPD3303X programmable power supply

The SPD3303X is a programmable high-precision three-channel linear power supply with a maximum output power of 220 watts.
SPECIFICATIONS
|
Number of channels |
CH1 : 0 ~ 32 V , 0 ~ 3.2 A |
|
Voltage range |
CH1: voltage: from 0 to 32V; current: from 0 to 3.2A |
|
Maximum output power |
220 W. |
|
Minimum voltage and current resolution |
1 mV, 1 mA |
|
Installation accuracy |
voltage: ± (0.03% + 10mV) current: ± (0.3% + 10mA) |
|
Display accuracy |
voltage: ± (0.03% + 10mV) current: ± (0.3% + 10mA) |
|
Voltage stabilization |
power supply effect: <0.01% + 3mV |
|
Current stabilization |
power effect: <0.2% + 3mA |
|
Display |
4.3 "TFT-LCD; bit size: 480 x 272 |
|
Bit size of the display |
5-bit voltage display; 4-bit current display |
|
Noise level |
≤1 mVrms (5 Hz - 1 MHz) |
|
Standard interfaces |
LAN, USB Device |
|
Power |
AC 100V / 120V / 220V / 230V ± 10%; 50/60 Hz |
|
Memory |
5 previous settings |
|
Dimensions |
225 × 143 × 278 mm |
Precision power supply OWON ODP3033
ODP3033 OWON - precision 3-channel programmable DC power supply (adjustable channels: 2x30V / 3A, + adjustable 6V / 3A, total power 198 W). Designed for DC power supply for the devices that are particularly sensitive to the quality of the supply voltage, it is characterized by low noise and high resolution in the installation of voltage and current.
SPECIFICATIONS
|
Number of channels |
2 (adjustable) |
1 (adjustable) |
||
|
Power outputs |
voltage |
from 0 to 30 V (independent / parallel) |
from 0 to 6 V |
|
|
current |
from 0 to 3 A (independent / sequential) |
from 0 to 3 A |
||
|
Power settings |
voltage |
≤0.01% + 3 mV |
≤0.01% + 3 mV |
|
|
current |
≤0.01% + 3 mA |
≤0.01% + 3 mA |
||
|
Load settings |
voltage |
≤0.01% + 3 mV |
≤0.01% + 3 mV |
|
|
current |
≤0.01% + 3 mA |
≤0.01% + 3 mA |
||
|
Ripples and noise (20 Hz - 20 MHz) |
voltage |
≤1 mV RMS. |
≤1 mV RMS. |
|
|
current |
≤5 mA RMS. |
≤4 mA RMS. |
||
|
Installation resolution |
voltage |
1 mV |
- |
|
|
current |
1 mA |
- |
||
|
Installation error |
voltage |
≤0.05% + 3 mV |
- |
|
|
current |
≤0.1% + 3 mA |
- |
||
|
Measurement resolution |
voltage |
1 mV (<10 V), 10 mV (> 10 V) |
- |
|
|
current |
1 mA |
- |
||
|
Measurement error |
voltage |
≤0.03% + 10 e.m.r. |
≤0.03% + 10 e.m.r. |
|
|
current |
≤0,1%+8 е.м.р. |
≤0,1%+5 е.м.р. |
||
|
Display |
TFT 4 ", 480 × 320, 65536 colors |
|||
|
Dimensions |
250 x 158 x 358 mm |
|||
XploRA PLUS Raman Spectrometer

Specifications
|
Parameters |
Value |
|
Accelerated mode of capturing SWIFT spectra |
yes, with a motorized table |
|
Confocal mapping |
0.5 μm XY |
|
Automation of basic operations |
OneClick procedure |
|
Optical microscope |
direct |
|
Spectral resolution on lattices 600, 1200, 1800, 2400, |
2… 8 cm-1 |
|
Wavelengths of excitation lasers |
532, 638, 785 nm |
|
Sample dimensions |
(5 ÷ 10 mm) x (5 ÷ 10 mm) x (0.2 ÷ 2 mm) |
|
drift AFM on XY |
2 nm / min |
IR Fourier spectrometer IRTracer-100

Specifications
|
Parameters |
Value |
|
Interferometer |
Michelson type with an angle of incidence of 300, equipped with an improved system dynamic adjustment, sealed (with indication of humidity |
|
Optical system |
One-beam |
|
Light dividers |
KBr plate with germanium coating for medium IR range |
|
Irradiation source |
Highly stable He-Ne laser; Highly stable high-energy ceramic source of medium and far IR range |
|
Detector |
Thermostabilized DLATGS for medium and far IR range |
|
Spectral range of wave numbers |
7800–350 cm-1 |
|
Accuracy of determination of wave numbers |
better than 0.01 cm-1 |
|
Resolution |
• 0.25; 0.5; 1; 2; 4; 8; 16 cm-1 (middle and far IR range) • 2; 4; 8; 16 cm-1 (close IR range) |
|
Signal to noise ratio |
60000: 1 and above (4 cm-1, 1 min, 2200 cm-1, peak to peak) |
|
The speed of movement of the mirror |
2; 2.8; 5; 9; 10; 20; 30 or 40 mm / min |
|
Ditch compartment |
200 x 230 x 170 mm, with automatic prefix recognition |
|
Software |
LabSolutions IR with a built-in library containing about 12,000 spectra of different classes of compounds |